Title
Inovativni funkcionalni čuvari prostora zasnovani na CAD/CAM tehnologiji
Creator
Nedeljković, Andrija, 1996-
CONOR:
57812233
Copyright date
2025
Object Links
Select license
Autorstvo-Bez prerade 3.0 Srbija (CC BY-ND 3.0)
License description
Dozvoljavate umnožavanje, distribuciju i javno saopštavanje dela, bez promena, preoblikovanja ili upotrebe dela u svom delu, ako se navede ime autora na način odredjen od strane autora ili davaoca licence. Ova licenca dozvoljava komercijalnu upotrebu dela. Osnovni opis Licence: http://creativecommons.org/licenses/by-nd/3.0/rs/deed.sr_LATN Sadržaj ugovora u celini: http://creativecommons.org/licenses/by-nd/3.0/rs/legalcode.sr-Latn
Language
Serbian
Cobiss-ID
Inventory ID
3861
Theses Type
Doktorska disertacija
description
Datum odbrane: 30.01.2025.
Other responsibilities
Academic Expertise
Medicinske nauke
University
Univerzitet u Kragujevcu
Faculty
Fakultet medicinskih nauka
Alternative title
Innovative fuctional space maintainers based on CAD/CAM technology
Publisher
[Andrija V. Nedeljković]
Format
71 list
Abstract (sr)
Uvod: S obzirom na to da pasivni fiksni čuvari prostora ne nadoknađuju izgubljeni zub
niti obezbeđuju funkciju žvakanja, cilj ovog istraživanja bio je da se ispita efikasnost
inovativnog pristupa očuvanju prostora nakon prevremenog vađenja prvih mlečnih
molara, primenom funkcionalnih čuvara prostora dizajniranih uz pomoć CAD/CAM
tehnologije.
Materijal i metode: Istraživanje je sprovedeno u vidu randomizovane kontrolisane
studije u koju je bilo uključeno ukupno 28 pacijenata, raspoređenih u dve grupe – 15 u
grupu sa čuvarima prostora izrađenim 3D štampom metala i 13 u grupu sa čuvarima
izrađenim glodanjem kompozitnog materijala. Kriterijumi za uključenje podrazumevali
su potrebu za ekstrakcijom ili postojeći gubitak prvog mlečnog molara usled
komplikacija karijesa, kao i predviđeno duže vreme do nicanja stalnog naslednika.
Funkcionalni čuvari prostora su izrađeni od različitih materijala u zavisnosti od
grupe: Grupa 3D štampe: legura metala, proizvedena od strane kompanije Scheftner iz
Nemačke. Grupa glodanja: visokofunkcionalni kompozit breCAM.HIPC (High Impact
Polymer Composite), proizveden od strane kompanije Bredent, takođe iz Nemačke.
Efikasnost čuvara prostora ocenjivana je na osnovu kliničkih pregleda nakon jednog, tri
i šest meseci od cementiranja, merenjem maksimalne sile zagrižaja, procenom
efikasnosti žvakanja i upitnikom zadovoljstva dece i roditelja, čime je dobijena
sveobuhvatna slika o funkcionalnosti ovih aparata.
Rezultati: Tokom šestomesečnog praćenja zabeležen je pozitivan klinički trend u obe
grupe, što ukazuje na adaptaciju pacijenata i stabilnost čuvara prostora. Ipak, analiza
je pokazala da tip čuvara prostora (3D štampani metalni u odnosu na glodani
kompozitni) značajno utiče na kliničke ocene (p < 0,001), što ukazuje na superiornost
određenih karakteristika u jednoj od grupa. Kada je reč o maksimalnoj sili zagrižaja, nije
bilo statistički značajne razlike između strane sa čuvarom prostora i kontralateralne
zdrave strane, ni u 3D štampanoj grupi (p = 0,270) ni u grupi sa glodanjem (p = 0,765), što
govori u prilog stabilnosti i funkcionalnosti oba tipa uređaja. Međutim, značajna
poboljšanja su uočena u oblasti efikasnosti žvakanja nakon postavljanja čuvara prostora,
što potvrđuje njihovu pozitivnu ulogu ne samo u mehaničkom očuvanju prostora, već i u
restauraciji osnovne fiziološke funkcije žvakanja. Analizom upitnika utvrđeno je da
je u obe grupe zabeležen visok stepen zadovoljstva. Deca su se brzo navikla na aparate,
dok su roditelji izrazili pozitivan stav u pogledu estetike, udobnosti i
funkcionalnosti, što dodatno potvrđuje kliničku primenljivost i prihvatljivost ovih
savremenih rešenj
Zaključak: Na osnovu šestomesečnog praćenja, može se zaključiti da funkcionalni
čuvari prostora izrađeni 3D štampom od metala pokazuju superiorne kliničke
karakteristike u odnosu na one izrađene glodanjem od kompozita. Iako razlike u
maksimalnoj sili zagrižaja i efikasnosti žvakanja nisu bile statistički značajne,
ukupni klinički ishodi ukazuju na veći stepen stabilnosti i adaptacije kod metalnih
čuvara prostora. Ovi nalazi podržavaju primenu CAD/CAM tehnologije kao novog
standardnog pristupa u izradi funkcionalnih čuvara prostora, koji ne samo da štite
zubni luk već aktivno doprinose rehabilitaciji oralne funkcije kod dece.
Abstract (en)
Introduction: Since passive fixed space maintainers do not replace the lost tooth nor restore
masticatory function, the aim of this study was to evaluate the effectiveness of an innovative
approach to space maintenance following premature extraction of the first primary molars,
through the use of functional space maintainers designed with CAD/CAM technology.
Materials and Methods: This study was conducted as a randomized controlled trial including a
total of 28 patients, divided into two groups – 15 patients received space maintainers fabricated
via 3D metal printing, while 13 received maintainers milled from composite material. The
inclusion criteria encompassed the need for extraction or the presence of first primary molar loss
due to caries complications, with a prolonged eruption period of the permanent successor. The
functional space maintainers were manufactured using different materials depending on the
group: 3D Printing Group: Metal alloy produced by Scheftner (Germany). Milling Group: Highperformance composite material breCAM.HIPC (High Impact Polymer Composite), produced by
Bredent (Germany). The effectiveness of space maintainers was assessed through clinical
examinations at one, three, and six months post-cementation, including measurements of
maximum bite force, evaluation of masticatory performance, and satisfaction questionnaires
completed by both children and parents. This comprehensive assessment provided valuable
insight into the functional success of these appliances.
Results: During the six-month follow-up period, a positive clinical trend was observed in both
groups, indicating successful patient adaptation and device stability. However, statistical analysis
showed that the type of space maintainer (3D-printed metal vs. milled composite) significantly
influenced clinical evaluations (p < 0.001), suggesting the superiority of certain features in one
group. Regarding maximum bite force, no statistically significant differences were found between
the side with the space maintainer and the contralateral healthy side in either the 3D printing group
(p = 0.270) or the milling group (p = 0.765), supporting the stability and functionality of both
types of appliances. Notably, significant improvements were observed in masticatory
performance following placement of the space maintainers, confirming their positive role not only
in maintaining the arch space mechanically, but also in restoring basic physiological chewing
function. Analysis of the satisfaction questionnaires revealed a high level of satisfaction in both
groups. Children adapted quickly to the devices, and parents expressed favorable opinions
regarding their aesthetic appeal, comfort, and functionality, further confirming the clinical
applicability and acceptability of these modern solutions.
Conclusion: Based on the six-month follow-up, it can be concluded that functional space
maintainers produced via 3D metal printing exhibit superior clinical performance compared to
those milled from composite. Although differences in maximum bite force and masticatory
performance were not statistically significant, the overall clinical outcomes indicate a greater
degree of stability and adaptation in the metal group. These findings support the use of CAD/CAM
technology as a new standard in the fabrication of functional space maintainers, which not only
preserve the dental arch but also actively contribute to the rehabilitation of oral function in
children.
Authors Key words
Čuvari prostora, CAD/CAM tehnologija, Glodanje, 3D štampa, Sila
zagrižaja, Efikasnost žvakanja
Authors Key words
Space maintainers, CAD/CAM technology, Milling, 3D printing, Bite force,
Masticatory performance
Classification
616.314-089.843(043.3)
Subject
Stomatologija, 3D štampa
Type
Tekst
Abstract (sr)
Uvod: S obzirom na to da pasivni fiksni čuvari prostora ne nadoknađuju izgubljeni zub
niti obezbeđuju funkciju žvakanja, cilj ovog istraživanja bio je da se ispita efikasnost
inovativnog pristupa očuvanju prostora nakon prevremenog vađenja prvih mlečnih
molara, primenom funkcionalnih čuvara prostora dizajniranih uz pomoć CAD/CAM
tehnologije.
Materijal i metode: Istraživanje je sprovedeno u vidu randomizovane kontrolisane
studije u koju je bilo uključeno ukupno 28 pacijenata, raspoređenih u dve grupe – 15 u
grupu sa čuvarima prostora izrađenim 3D štampom metala i 13 u grupu sa čuvarima
izrađenim glodanjem kompozitnog materijala. Kriterijumi za uključenje podrazumevali
su potrebu za ekstrakcijom ili postojeći gubitak prvog mlečnog molara usled
komplikacija karijesa, kao i predviđeno duže vreme do nicanja stalnog naslednika.
Funkcionalni čuvari prostora su izrađeni od različitih materijala u zavisnosti od
grupe: Grupa 3D štampe: legura metala, proizvedena od strane kompanije Scheftner iz
Nemačke. Grupa glodanja: visokofunkcionalni kompozit breCAM.HIPC (High Impact
Polymer Composite), proizveden od strane kompanije Bredent, takođe iz Nemačke.
Efikasnost čuvara prostora ocenjivana je na osnovu kliničkih pregleda nakon jednog, tri
i šest meseci od cementiranja, merenjem maksimalne sile zagrižaja, procenom
efikasnosti žvakanja i upitnikom zadovoljstva dece i roditelja, čime je dobijena
sveobuhvatna slika o funkcionalnosti ovih aparata.
Rezultati: Tokom šestomesečnog praćenja zabeležen je pozitivan klinički trend u obe
grupe, što ukazuje na adaptaciju pacijenata i stabilnost čuvara prostora. Ipak, analiza
je pokazala da tip čuvara prostora (3D štampani metalni u odnosu na glodani
kompozitni) značajno utiče na kliničke ocene (p < 0,001), što ukazuje na superiornost
određenih karakteristika u jednoj od grupa. Kada je reč o maksimalnoj sili zagrižaja, nije
bilo statistički značajne razlike između strane sa čuvarom prostora i kontralateralne
zdrave strane, ni u 3D štampanoj grupi (p = 0,270) ni u grupi sa glodanjem (p = 0,765), što
govori u prilog stabilnosti i funkcionalnosti oba tipa uređaja. Međutim, značajna
poboljšanja su uočena u oblasti efikasnosti žvakanja nakon postavljanja čuvara prostora,
što potvrđuje njihovu pozitivnu ulogu ne samo u mehaničkom očuvanju prostora, već i u
restauraciji osnovne fiziološke funkcije žvakanja. Analizom upitnika utvrđeno je da
je u obe grupe zabeležen visok stepen zadovoljstva. Deca su se brzo navikla na aparate,
dok su roditelji izrazili pozitivan stav u pogledu estetike, udobnosti i
funkcionalnosti, što dodatno potvrđuje kliničku primenljivost i prihvatljivost ovih
savremenih rešenj
Zaključak: Na osnovu šestomesečnog praćenja, može se zaključiti da funkcionalni
čuvari prostora izrađeni 3D štampom od metala pokazuju superiorne kliničke
karakteristike u odnosu na one izrađene glodanjem od kompozita. Iako razlike u
maksimalnoj sili zagrižaja i efikasnosti žvakanja nisu bile statistički značajne,
ukupni klinički ishodi ukazuju na veći stepen stabilnosti i adaptacije kod metalnih
čuvara prostora. Ovi nalazi podržavaju primenu CAD/CAM tehnologije kao novog
standardnog pristupa u izradi funkcionalnih čuvara prostora, koji ne samo da štite
zubni luk već aktivno doprinose rehabilitaciji oralne funkcije kod dece.
“Data exchange” service offers individual users metadata transfer in several different formats. Citation formats are offered for transfers in texts as for the transfer into internet pages. Citation formats include permanent links that guarantee access to cited sources. For use are commonly structured metadata schemes : Dublin Core xml and ETUB-MS xml, local adaptation of international ETD-MS scheme intended for use in academic documents.